Abstract
Background
Access to left atrial appendage closure (LAAC) in Canada is limited due to funding restrictions. This work aimed to assess Canadian clinical practice on patient selection, post-procedural antithrombotic therapy, and safety and/or efficacy with WATCHMAN implantation.
Methods
Seven Canadian centers implanting WATCHMAN participated in this prospective multicenter, observational registry. All procedures were done under general anesthesia with transesophageal echocardiography guidance. Patients were prospectively followed for 2years. Long-term stroke rate was compared with the expected rate based on CHA2DS2-VASc score.
Results
A total of 272 patients who underwent WATCHMAN LAAC between December 2013 and August 2019 (mean age [SD]: 75.4 [8.75] years; male, 63.2%; CHA2DS2-VASc: 4.35 [1.64]; HAS-BLED: 3.55 [0.94]) were included. Most patients (90.4%) had prior history of bleeding (80.5% major, 21.7% minor). The WATCHMAN device was successfully implanted in 269 patients (98.9%), with a few procedure-related complications including 5 pericardial effusions requiring drainage (1.8%) and 1 death (0.4%; 22 days post-LAAC from respiratory failure). Post-LAAC antithrombotic therapy included dual antiplatelet therapy in 70.6%, single antiplatelet therapy 18.4%, and OAC 13.6%. During the follow-up period (mean 709.7 [467.2] days), there was 81.4% reduction of ischemic stroke rate based on the expected rate from CHA2DS2-VASc score (6.0% expected vs. 1.1% observed). Device-related thrombus was detected in 1.8%.
Conclusion
The majority of Canadian patients who underwent LAAC had OAC contraindication due to prior bleeding, and most were safely treated with antiplatelet therapy post-LAAC with low device-related thrombus incidence. Long-term follow-up demonstrated that LAAC achieved a significant reduction in ischemic stroke rate.
Introduction
Atrial fibrillation (AF) is the most common cardiac arrhythmia and a major cause of stroke, responsible for 15% of all strokes and 30% of strokes in patients age>80.(
1- Rosamond W.
- Flegal K.
- Furie K.
- et al.
Heart disease and stroke statistics--2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee.
) Importantly, strokes associated with AF are more severe; AF-related stroke victims have 50% greater likelihood to become disabled, and >50% likelihood of death.(
2- Lamassa M.
- Di Carlo A.
- Pracucci G.
- et al.
Characteristics, outcome, and care of stroke associated with atrial fibrillation in Europe: data from a multicenter multinational hospital-based registry (The European Community Stroke Project).
,
3- Skanes A.C.
- Healey J.S.
- Cairns J.A.
- et al.
Focused 2012 update of the Canadian Cardiovascular Society atrial fibrillation guidelines: recommendations for stroke prevention and rate/rhythm control.
) Accordingly, stroke prevention is one of the main pillars of AF management; the Canadian Cardiovascular Society recommends anticoagulation for most patients with age 65 years or older or CHADS
2 score ≥ 1 and the European Society of Cardiology recommends anticoagulation for CHA
2DS
2-Vasc≥1.(
4- Andrade J.G.
- Aguilar M.
- Atzema C.
- et al.
The 2020 Canadian Cardiovascular Society/Canadian Heart Rhythm Society Comprehensive Guidelines for the Management of Atrial Fibrillation.
,
5- Hindricks G.
- Potpara T.
- Dagres N.
- et al.
2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC.
)
The benefit of OAC in stroke prevention has to be balanced with the risk of major bleeding. Despite the safety profile of non-vitamin K oral anticoagulation (NOAC), the annual incidence of major bleeding ranges from 2.13% to 3.6%, with an annual incidence of intracranial hemorrhage ranging from 0.1% to 0.5%.(
6- Connolly S.J.
- Ezekowitz M.D.
- Yusuf S.
- et al.
Dabigatran versus warfarin in patients with atrial fibrillation.
,
7- Patel M.R.
- Mahaffey K.W.
- Garg J.
- et al.
Rivaroxaban versus warfarin in nonvalvular atrial fibrillation.
,
8- Granger C.B.
- Alexander J.H.
- McMurray J.J.
- et al.
Apixaban versus warfarin in patients with atrial fibrillation.
) Therefore, even though OAC is effective for thromboembolic prevention, there remain a large proportion of eligible patients not on therapy for multiple reasons. These challenges have led to the investigations of device-based therapies for non-valvular AF including percutaneous left atrial appendage closure (LAAC) to prevent stroke, which is a major source of thrombus related to AF. Randomized controlled trials (PROTECT-AF & PREVAIL) have shown safety and efficacy of the WATCHMAN device in comparison to warfarin in patients eligible for OAC.(
9- Holmes D.R.
- Reddy V.Y.
- Turi Z.G.
- et al.
Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial.
,
10- Holmes Jr., D.R.
- Kar S.
- Price M.J.
- et al.
Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial.
,
11- Reddy V.Y.
- Doshi S.K.
- Kar S.
- et al.
5-Year Outcomes After Left Atrial Appendage Closure: From the PREVAIL and PROTECT AF Trials.
) Among patients with contraindications to OAC, randomized trials comparing LAAC to antiplatelet/no therapy are ongoing, but enrolment rates have been very slow, and the ASAP-TOO trial was stopped prematurely due to recruitment challenge.(
12- Holmes D.R.
- Reddy V.Y.
- Buchbinder M.
- et al.
The Assessment of the Watchman Device in Patients Unsuitable for Oral Anticoagulation (ASAP-TOO) trial.
) Nevertheless, based on cumulative evidence, LAAC is given a class IIB recommendation for patients with high stroke-risk and contraindications to long-term OAC.(
5- Hindricks G.
- Potpara T.
- Dagres N.
- et al.
2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC.
)
Despite accumulating data, access to LAAC in Canada is limited due to funding restrictions. We aimed to assess the early Canadian clinical experience with LAAC, assessing patient selection, procedural outcomes, post-procedural antithrombotic therapy, and the safety and long-term efficacy with WATCHMAN implantation.
Methods
The Canadian WATCHMAN Registry is a multicenter prospective, non-randomized, observational registry enrolling consecutive patients undergoing LAAC with the WATCHMAN legacy device in Canadian cardiac catheterization or electrophysiological laboratories. All Canadian centers implanting WATCHMAN were invited to participate in this multicenter registry; 7 of 11 sites participated. Institutional research ethics boards approvals were obtained and patients gave informed consents for prospective follow-up.
Patient Population
Patients with non-valvular AF (paroxysmal, persistent or permanent) who met the following inclusion and exclusion criteria were included. Inclusion criteria included 1) age ≥18 years, and 2) CHADS2 ≥1 and/or CHA2DS2-Vasc ≥2, and 3) prior major bleeding (intracranial, gastrointestinal bleeding, intraocular, respiratory, genitourinary, retroperitoneal, pericardial, anemia requiring transfusions, etc.), or contraindications to long-term OAC (HASBLED ≥3, high fall risk, cerebral aneurysm, blood dyscrasias, aortic dissection, renal failure, etc.), or failure of OAC (stroke/TIA while on OAC), or patients deemed not suitable for long-term anticoagulation. Patients with the presence of LAA thrombus or severe untreated mitral stenosis were excluded.
Pre-procedure imaging
Pre-procedural baseline transesopageal echocardiography (TEE) was routinely recommended, or alternatively, cardiac computed tomography angiography (CCTA) was performed instead of TEE pre-procedure, to rule out pre-existing LAA thrombus and to evaluate LAA anatomy and dimensions for accurate measurement of the widest LAA ostium (usually at the level of the circumflex artery) at 0, 45, 90, and 135 degrees, and also the available depth of the LAA (from ostium to apex of LAA).
Procedural details
All procedures were done under general anesthesia with TEE guidance. Intravenous heparin was administered pre- or immediately following transseptal puncture to maintain activated clotting time >250-300s during the procedure. To ensure adequate mean left atrial pressure (>12mmHg) for more accurate LAA measurements, saline bolus was administered if necessary. Device sizing selection was based upon the maximum LAA ostium diameter. Oversizing was recommended by 9-25% based on the widest LAA measurement. Before device release, fulfillment of 4 criteria (PASS) was confirmed: (
1- Rosamond W.
- Flegal K.
- Furie K.
- et al.
Heart disease and stroke statistics--2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee.
) Position (device distal or at the ostium of the LAA; protrusion of shoulder by <5-7mm was acceptable), (
2- Lamassa M.
- Di Carlo A.
- Pracucci G.
- et al.
Characteristics, outcome, and care of stroke associated with atrial fibrillation in Europe: data from a multicenter multinational hospital-based registry (The European Community Stroke Project).
) Anchor (testing stability by retracting the deployment knob and letting go, to assess return to original position), (
3- Skanes A.C.
- Healey J.S.
- Cairns J.A.
- et al.
Focused 2012 update of the Canadian Cardiovascular Society atrial fibrillation guidelines: recommendations for stroke prevention and rate/rhythm control.
) Size (device shoulder compressed 8-20% of original size on TEE) (
table 2), and (
4- Andrade J.G.
- Aguilar M.
- Atzema C.
- et al.
The 2020 Canadian Cardiovascular Society/Canadian Heart Rhythm Society Comprehensive Guidelines for the Management of Atrial Fibrillation.
) Seal (assess TEE for any residual flow; must be <5mm before release). When all criteria were met, the device was released. Final angiography and TEE assessment were then performed.
Table 2Procedural Details
Abbreviations: SD, standardized difference; TEE, transesophageal echocardiography; LAA, left atrial appendage; CTA, computed tomography angiography
Post-procedure dual antiplatelet therapy (DAPT) with aspirin 81mg/d and clopidogrel 75mg/d is commonly recommended for 3 months and followed by life-long aspirin alone; however, the post-LAAC antithrombotic regimen was at physician discretion. Post-procedural imaging with TEE or CCTA were recommended 1-6 months post-LAA closure.
Clinical follow-up
Long-term events (stroke, transient ischemic attack [TIA], systemic embolism [SE], cardiovascular [CV] death, non-CV death, major bleeding, and minor bleeding) were collected prospectively. Clinical or telephone follow-up were obtained at 3, 12 and 24 months post-LAAC.
Definitions
Procedural major adverse events (MAE) were defined as a composite of death, device embolization, stroke, SE, myocardial infarction, cardiac tamponade, and major bleeding. Other procedural complications (pericardial effusion not requiring drainage and minor bleeding) were also assessed. For long-term follow-up, major adverse cardiovascular events (MACE), defined as a composite of death, stroke, TIA, SE, and MI, were assessed. In addition, major and minor bleeding was also evaluated. For patients who underwent device surveillance (CCTA or TEE), the incidence of device-related thrombus (DRT) and peri-device leak (PDL) was assessed. The definition of DRT and PDL was in accordance with previous studies.(
13- Saw J.
- Tzikas A.
- Shakir S.
- et al.
Incidence and Clinical Impact of Device-Associated Thrombus and Peri-Device Leak Following Left Atrial Appendage Closure With the Amplatzer Cardiac Plug.
)
Statistical analysis
For continuous variables, means and standard deviations were calculated and compared using the Wilcoxon test. Binary variables are reported as counts and percentages and between-group differences were assessed using the χ2 test. The efficacy of WATCHMAN in preventing stroke, TIA, and systemic embolism was tested by comparing the observed event-rate at follow-up with the predicted event-rate by the CHADS
2 and CHA
2DS
2-Vasc scores (
1- Rosamond W.
- Flegal K.
- Furie K.
- et al.
Heart disease and stroke statistics--2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee.
). The average annual risk for the whole study population was calculated from the predicted individual patient annual risk. The observed annualized ischemic stroke rate and thromboembolic event-rate (stroke, TIA, and SE) were subtracted from the predicted event-rates, and divided by the predicted event-rate x100, to obtain the % relative risk reduction (% relative reduction).
Data were analyzed using R software version 4.0.5 (R Foundation for Statistical Computing, Vienna, Austria). All p values were 2-sided, and significance was defined as p<0.05 for all analyses.
Discussion
In this largest Canadian prospectively registry of patients undergoing LAAC with the WATCHMAN device to-date, we found that (a) >1/3 of patients had a history of cerebrovascular disease and >90% had a prior history of bleeding events, (b) the WATCHMAN device was successfully implanted in the majority of cases (98.9%) with 3.3% peri-procedural MAE including 5 pericardial effusions requiring drainage (1.8%) and 1 death due to respiratory failure (0.4%), (c) DAPT was the predominant antithrombotic regimen after LAAC (70.6%) and OAC was used only in 18.4%, and (d) the follow-up data showed a 81.4% reduction of ischemic stroke rate based on the expected rate from CHA2DS2-VASc score (6.0% expected vs. 1.1% observed).
This study cohort comprises the early Canadian experience with the WATCHMAN legacy device. Health Canada approved the WATCHMAN device in Jan 2016, and prior to this time period, implant of WATCHMAN was under the special access program. Thus, patient selection for this procedure was rigorous and restricted, and a high proportion of our patients had prior bleeding events (90.4%), with most having had major bleeding (>1/3 had prior intracranial bleeding). The National Cardiovascular Data Registry LAAO Registry, a prospective, nationwide registry designed to function as the formal post-market surveillance in the US, enrolled patients undergoing LAAC with WATCHMAN during a similar timeframe (between January 2016 and December 2018), and showed lower proportions of prior major bleeding and intracranial bleeding (69.5% vs. 80.5%, and 11.9% vs. 36.4%, respectively) compared to our Canadian cohort.(
14- Price M.J.
- Slotwiner D.
- Du C.
- et al.
Clinical Outcomes at 1 Year Following Transcatheter Left Atrial Appendage Occlusion in the United States.
) This disparity is largely due to the stricter indication for LAAC in Canada due to funding restrictions, where percutaneous LAAC is generally performed in patients with high stroke-risk and absolute/relative contraindications to OAC or failure on OAC. There is yet to be published randomized trial data on LAAC in patients contraindicated to OAC. The ASAP-TOO study was stopped prematurely,(
15- Marks D.S.
- Mensah G.A.
- Kennard E.D.
- Detre K.
- Holmes Jr., D.R.
Race, baseline characteristics, and clinical outcomes after coronary intervention: The New Approaches in Coronary Interventions (NACI) registry.
) while the STROKE-CLOSE (NCT02830152) and CLOSURE-AF (NCT03463317) studies are ongoing but enrolment rates are slow. Thus, real-world registries evaluating the effectiveness and safety of LAAC in patients contraindicated to OAC remain highly relevant in the current era. There are also ongoing large, randomized trials comparing LAAC versus NOACs in AF patients suitable for OAC (e.g. the CHAMPION-AF (NCT04394546) and CATALYST (NCT04226547) trials), which may expand the indications for LAAC in Canada in the near future.
The technical success rate with this device in our study was 98.9%, which was higher than that of pivotal trials such as PROTECT-AF and PREVAIL,(
9- Holmes D.R.
- Reddy V.Y.
- Turi Z.G.
- et al.
Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial.
,
10- Holmes Jr., D.R.
- Kar S.
- Price M.J.
- et al.
Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial.
) and compatible with recent registries.(
16- Boersma L.V.
- Schmidt B.
- Betts T.R.
- et al.
Implant success and safety of left atrial appendage closure with the WATCHMAN device: peri-procedural outcomes from the EWOLUTION registry.
,
17- Freeman J.V.
- Varosy P.
- Price M.J.
- et al.
The NCDR Left Atrial Appendage Occlusion Registry.
) The rate of in-hospital adverse events in our registry was also compatible with that of pivotal trials and other registries, (
9- Holmes D.R.
- Reddy V.Y.
- Turi Z.G.
- et al.
Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial.
,
10- Holmes Jr., D.R.
- Kar S.
- Price M.J.
- et al.
Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial.
,
17- Freeman J.V.
- Varosy P.
- Price M.J.
- et al.
The NCDR Left Atrial Appendage Occlusion Registry.
,
18- Holmes Jr., D.R.
- Reddy V.Y.
- Gordon N.T.
- et al.
Long-Term Safety and Efficacy in Continued Access Left Atrial Appendage Closure Registries.
) with the most frequent peri-procedure complications being cardiac tamponade requiring drainage. The improved implant success rate of recent registries including our study is attributed to increasing operator experience and likely complimented by the use of pre-procedure CT assessment. A learning curve phenomenon was previously described with LAAC, and 30 cases was proposed as a threshold to reach proficiency and to optimize clinical outcomes.(
19- Jung R.G.
- Simard T.
- Killu A.
- et al.
Learning Curve and Outcomes of Left Atrial Appendage Closure.
) Similar to other structural heart disease interventions, CT assessment has been widely used as a pre-procedure evaluation and can provide accurate LAA morphology and sizing.(
20- Saw J.
- Fahmy P.
- Spencer R.
- et al.
Comparing Measurements of CT Angiography, TEE, and Fluoroscopy of the Left Atrial Appendage for Percutaneous Closure.
) The next-generation WATCHMAN FLX device is now commercially available, and early studies have shown this new device to be associated with a lower incidence of adverse events and high incidence of effective appendage closure.(
21- Kar S.
- Doshi S.K.
- Sadhu A.
- et al.
Primary Outcome Evaluation of a Next-Generation Left Atrial Appendage Closure Device: Results From the PINNACLE FLX Trial.
)
In the Canadian WATCHMAN Registry, the majority of patients were managed with antiplatelet therapy rather than anticoagulation therapy. In most patients (>70%), DAPT was prescribed and de-escalated to single antiplatelet therapy (SAPT) at the time of 1-6 months after the procedure. In selected patients with a very high-risk for bleeding (∼10%), SAPT was the initial antithrombotic regimen post-LAAC. In contrast, in the early US experience with the WATCHMAN legacy device, >90% of patients undergoing LAAC were discharged home on OACs with either warfarin or NOAC.(
14- Price M.J.
- Slotwiner D.
- Du C.
- et al.
Clinical Outcomes at 1 Year Following Transcatheter Left Atrial Appendage Occlusion in the United States.
) Despite the significant difference in antithrombotic regimen, the incidence of ischemic stroke in our cohort was 3.7% during the mean 2yr follow-up, which was comparable to that in the US cohort (1.4% during 1-year follow-up). Importantly, the rate of major bleeding in our study cohort was lower than that in the US cohort (8.8% for mean 2-year follow-up vs. 7.9% for 1-year follow-up).(
14- Price M.J.
- Slotwiner D.
- Du C.
- et al.
Clinical Outcomes at 1 Year Following Transcatheter Left Atrial Appendage Occlusion in the United States.
) In addition, >80% of patients underwent post-implant surveillance using either CT or TEE, and DRT was detected only in 1.8%, which appeared lower than previously reported (3.8%).(
22- Saw J.
- Nielsen-Kudsk J.E.
- Bergmann M.
- et al.
Antithrombotic Therapy and Device-Related Thrombosis Following Endovascular Left Atrial Appendage Closure.
) Given the lower bleeding event with similar efficacy to prevent ischemic events and DRT, our data suggests that an antiplatelet-dominant antithrombotic strategy post-LAAC is safe. Further investigation is warranted to assess the optimal antithrombotic regimen.
Limitations
There are several limitations in our study. Our Canadian WATCHMAN Registry included 7 of 11 sites performing LAAC in Canada, thus, we did not enroll all WATCHMAN cases performed in Canada during the time period. Nevertheless, we believe our registry was representative of the clinical practice, patient selection, and outcomes of real-world LAAC in Canada. This is an observational registry; thus the use of pre-procedural imaging, device surveillance post-procedure, and use of antithrombotic regimen was at the discretion of the physicians. Although the clinical data were prospectively collected by each center, the events and imaging results were not adjudicated. The new-generation WATCHMAN FLX device has now largely replaced the WATCHMAN legacy device. Given the excellent performance of WATCHMAN FLX as reported in the PINNACLE Registry,(
21- Kar S.
- Doshi S.K.
- Sadhu A.
- et al.
Primary Outcome Evaluation of a Next-Generation Left Atrial Appendage Closure Device: Results From the PINNACLE FLX Trial.
) its implementation may influence safety and effectiveness, but this new device was not evaluated in our cohort.
Article info
Publication history
Accepted:
March 21,
2023
Received:
February 27,
2023
Publication stage
In Press Journal Pre-ProofFootnotes
Disclosures
This study is supported by unrestricted research grant from Boston Scientific. Dr. Saw has received unrestricted research grant supports (from the Canadian Institutes of Health Research, Heart & Stroke Foundation of Canada, National Institutes of Health, University of British Columbia Division of Cardiology, AstraZeneca, Abbott, St Jude Medical, Boston Scientific, and Servier), salary support (Michael Smith Foundation of Health Research), speaker honoraria (AstraZeneca, Abbott, Boston Scientific, and Sunovion), consultancy and advisory board honoraria (AstraZeneca, St Jude Medical, Abbott, Boston Scientific, Baylis, Gore, FEops), and proctorship honoraria (Abbott, St Jude Medical and Boston Scientific). Dr. Inohara has received speaker honoraria (Boston Scientific and Edwards Lifesciences). Dr. Gilhofer has no disclosures. Dr. Morillo has received consultancy, advisory board and speaker honoraria from; Abbott, Boston Scientific, Medtronic, Servier, Novartis, and BMS. Dr. Ibrahim received research support from Biomodex, Edwards, Medtronic, Opsens; speaker honoraria from Boston Scientific, Edwards, Medtronic, Opsens; and proctor for Abbott and Edwards. Dr. Pearce received speaker honoraria from BMS, Servier, Pfizer, Boston Scientific, Bayer and Boerhinger.
Copyright
© 2023 Published by Elsevier Inc. on behalf of the Canadian Cardiovascular Society.